Bedside primer

Principles of Impella

Everything you need to finish the lab cases: what a microaxial pump actually does, what SmartAssist shows, how to titrate P-levels, and how to recognize suction, malposition, afterload limit, purge failure, RV unmasking, and a token ECMELLA — from the pattern, not from a number the AIC does not calculate.

A microaxial pump, not a balloon

Impella is a catheter-mounted microaxial pump that pulls blood from the ventricle and ejects it into the great vessel. For the LV devices (CP, 5.5) that means LV → aorta across the aortic valve. For RP it means IVC/RA → PA. It unloads the chamber it sits in and forwards stroke volume. It is not IABP counterpulsation and it is not VA-ECMO.

  • Impella CP — 14-Fr femoral percutaneous. Peak flow about 3.5–4 L/min in a filled ventricle.
  • Impella 5.5 — 19-Fr, usually a surgical axillary graft. Peak about 5.5–6.2 L/min. Same SmartAssist rules, hungrier inlet.
  • Impella RP — femoral venous, inlet in the IVC/RA, outlet in the PA. Forwards a failed RV.
  • ECMELLA — VA-ECMO plus an LV Impella. ECMO is the afterload; Impella is the unload.

Abiomed / J&J IFU language is consistent across platforms: confirm position, then titrate. The motor does not invent blood.

What SmartAssist actually shows

The Automated Impella Controller (AIC) with SmartAssist is the screen you will live on. Three traces matter: a red Ao placement signal, a white LV estimate, and a green motor current. Displayed flow is derived from current and P-level. SmartAssist can also trend CO and CPO; it still does not print a diagnosis. How to read the traces is the next section — it is the skill that separates a titration from a guess.

ReadoutWhat it isHow you use it
P-levelThe support setting you choose (P1–P9)Not RPM. Titrate after position is confirmed.
Flow (Q)Estimated L/minFalls with suction, afterload, malposition. Not a promise.
Placement signalAo (red) + LV (white)Morphology tells you which side of the valve the sensor is on.
Motor currentElectrical load, greenPulsatile = two chambers. Flat = same chamber or no native beat.
Purge pressureCassette line pressureRoughly 300–700 mmHg. Climbing is a cassette problem.
Purge flowmL/h of D5W ± heparinDribble plus high pressure = fix the purge.

Reading the AIC waveforms

The Placement screen is two stacked windows. The lab console is drawn to match the real AIC. Learn the pictures here, then open a case and name the tracing before you touch P-level. Waveforms are how you decide whether the next click is support, volume, a pull-back, a pressor wean — or leaving the catheter alone.

What the screen is for

SmartAssist is not a diagnosis engine. It is three synchronized pictures of the same pump: where the optical sensor sits, what the estimated LV is doing, and how hard the motor is working to move blood from one chamber to the other. You use them, in that order, to answer four questions:

  1. Is the catheter across the aortic valve?
  2. Is the inlet seeing blood, or sucking wall / apparatus / vacuum?
  3. Is the motor loaded (afterload, clot) or unloaded (no work, no spin)?
  4. Is there a native beat left, or is the pump the entire circulation?

Displayed flow is derived from motor current and P-level. It will lie if position is wrong, if the inlet is empty, or if the motor is no longer moving blood. The traces are the ground truth. The number is a guess.

The three traces

  • Red Ao (placement signal) — optical sensor at the outlet, where the cannula meets the outflow. When the device is seated this is an aortic pressure waveform: upstroke, dicrotic notch, diastolic decay. It is not your arterial line. Manage blood pressure off a real A-line. The red trace answers “which side of the valve is the sensor on?”
  • White LV (SmartAssist estimate) — not a second transducer. It is derived from the Ao signal plus motor current. When seated it should look ventricular: high systolic peak, deep diastolic valley. That valley is the visual signature that the inlet is in the LV. A white trace that never dives is either the wrong chamber or an LV that is still loaded (token ECMELLA, no native beat, afterload clamp).
  • Green motor current — energy the motor is using to move blood, in milliamps. When inlet and outlet sit in two different pressure chambers, current is pulsatile: it peaks when the aortic valve is open (smaller LV–Ao gradient, more flow through the cannula) and falls when the valve is closed. Flat current means both ends of the pump see the same chamber — or the ventricle is not contracting. Mean current rising with flow stuck is afterload. Mean current falling with flow collapsing is a motor that has stopped doing work.

How the traces line up on one beat

Read them as one cardiac cycle, not three independent graphs. On a seated device:

  • Systole — white LV peaks first. Red Ao upstroke follows. Green motor current rises (easier pump, more flow).
  • Early diastole — red notch, then decay. White falls toward its valley. Green current falls.
  • Late diastole — this is where suction declares. A healthy white valley stays a few mmHg above zero and recovers before the next beat. A suction valley crashes below zero. Continuous suction never recovers.

If the three traces no longer share a rhythm, believe arrhythmia or a placement signal the AIC has already flagged as unreliable — not a new P-level.

Does the LV waveform ever cross the Ao waveform?

This is the native-ejection test. On a seated device, look at whether the white LV peak climbs through the red Ao line each beat.

  • White crosses red — the LV generated a higher pressure than the aorta. The aortic valve opened. There is leftover native stroke volume on top of pump flow. Red keeps an aortic pulse and a notch; green motor current is clearly pulsatile.
  • White never crosses red — the LV never generates enough pressure to open the aortic valve. Native cardiac output contributes little (often ~0) to overall cardiac output. Displayed Impella flow is the cardiac output. Red collapses to a near-flat line (pulse pressure of a few mmHg). White stays ventricular — it still dives in diastole — but its systolic peak remains below the Ao line. Green loses most of its pulse because there is no AV opening to cycle the load; mean current stays in the working band if the pump is actually moving blood.

This is not malposition. Red is still aortic (flat, but aortic). White is still ventricular. The inlet is doing its job — sometimes so well, at P-8 / P-9 on a 5.5, that the ventricle is fully unloaded and the valve never opens. Echo will show a closed AV and an inlet mid-cavity. Do not pull that catheter.

Split it from token ECMELLA: same “AV closed” physics, opposite LV. Token VA afterload leaves white diastolic high (the cavity is still loaded). A working Impella that has taken over the circulation leaves white diastolic low or slightly negative, with high displayed flow and a CO that almost equals pump flow.

The numbers next to the traces

BoxWhat it printsWhat you actually use
Ao 88/72 (76)Sys / dias (mean), mmHgMorphology first. Mean is not an A-line MAP.
LV 90/12Estimated LV sys / diasDiastole is the suction tell. Negative = empty inlet.
MC 822/738 (762)Max / min (mean), mAPulse of current = pulsatility. Mean rising = load. Mean falling = no work.
Flow 3.8 / 3.0 → 3.3Max / min → mean L/minMax–min shrinks in suction. Mean stuck low = afterload or position.

Motor-current pulse pressure (max − min) is the fastest “is it across the valve?” check. Seated CP at working P is typically tens to a couple of hundred mA of pulse. Same-chamber or no-native-beat tracings sit under ~20 mA. Afterload keeps the pulse and lifts the mean. Motor failure drops the mean.

One algorithm, every time

  1. Is motor current pulsatile? Yes → the pump is crossing a valve (or you have a beating ventricle). No → either both ends are on the same side of the AV, or there is no native contraction. You cannot tell those two apart from the green trace alone. Echo.
  2. Do Ao and LV look like different chambers? Red aortic + white ventricular = seated. Both ventricular = too deep (sensor in the LV). Both aortic = too shallow (sensor in the aorta).
  3. Does white ever cross red? If the LV peak stays below the Ao line, the aortic valve is not opening. Native output is negligible. Displayed pump flow is the cardiac output. Echo before you call malposition.
  4. Look at white diastole. A deep negative valley that recovers by end-diastole is diastolic suction (empty LV). A valley that never recovers is continuous suction (empty, obstructed, or against the wall).
  5. Look at green mean, not just shape. High mean + low displayed flow = afterload or a working motor against a brick wall. Low mean + low flow at the same P = the motor is no longer doing work (clot, seized purge, inlet occluded).
  6. Then look at the patient. Suction plus a high CVP and a low PAPi is the RV, not a dry tank. High MAP, high motor current, low displayed flow is afterload, not under-P. Equalized filling pressures and a swinging septum is tamponade, not “needs volume.”

Atlas at a glance

Same three traces. The diagnosis is the combination, not any one line.

PatternRed AoWhite LVGreen MCFirst move
SeatedAortic, notchVentricular valleyPulsatile, mid meanTitrate P
Too deep (in LV)VentricularVentricularFlatP-2, pull 1 cm
Too shallow (in Ao)AorticAorticFlatP-2, advance 1 cm
HypovolemiaAorticDiastole < 0, recoversPulses, diastolic chatterEase P, fill, find the leak
Continuous suctionOverlaying whiteStays negativeFlat + chatterDrop P now, echo
RV failureLooks like hypovolemiaEmpty LV valleySame as suctionEase P, treat the RV
TamponadeLow pulseSuction-likeLow pulseDrain. Do not fill blindly.
High afterloadTall aorticValley presentHigh mean, pulse keptWean pressor / vasodilate
No native ejectionFlat aorticVentricular, never crosses redSmall ripple, mid meanEcho. Do not pull.
VTFast spikesFast spikesIrregularTreat the rhythm
Motor failingDampedDamped, diastole highMean collapsedEcho, prepare exchange
Token ECMELLAPP ~4 mmHgDiastolic highBarely pulsesTitrate unload, not P-2
RP seatedPA pulseCVP lowerPulsatileTitrate RP
RP failingPA pulse goneCVP high / overlayFlatFix the RV forward

What the screen should look like

Each card is the Placement window (red Ao + white LV, or PA + CVP on RP) over the motor-current window (green). Same colors and sweep as the AIC in the lab. Name the tracing, then read why.

Correct positionseated
160
-40
Ao (mmHg)
88/72(76)
LV
90/12
1022
502
Motor Current (mA)
822/738(762)
10 sec.
  • Red aortic
  • White ventricular
  • Green pulsatile
Red is aortic (notch, modest pulse). White is ventricular (dives in diastole). Green is pulsatile, mean in the working band (~500–900 mA on a filled CP). Inlet mid-LV, outlet above the AV. This is the only tracing you titrate P on. Save it as the patient’s baseline — every later alarm is a departure from this picture.
Too deep — Impella in the ventriclein lv
Impella Position in Ventricle
160
-40
Ao (mmHg)
94/12(39)
LV
96/4
1012
492
Motor Current (mA)
756/749(752)
10 sec.
  • Both ventricular
  • Green flat
Both traces look ventricular. Motor current is a flat line — inlet and outlet are in the same chamber, so there is no cyclic load. Displayed flow will lie (the algorithm still prints a number). The pigtail and inlet have gone too far toward the apex or the mitral apparatus. Drop to P-2. Pull back 1 cm at a time under echo or fluoro until the red trace becomes aortic and the green trace pulses again. Climbing P here shears cells against the MV.
Too shallow — Impella in the aortain ao
Impella Position in Aorta
160
-40
Ao (mmHg)
86/68(74)
LV
84/64
866
346
Motor Current (mA)
610/602(606)
10 sec.
  • Both aortic
  • Green flat
Both traces look aortic (or the white estimate collapses onto the red). Motor current is flat again — same-chamber physics. The pigtail and inlet have come back across the valve; the pump is circulating aortic blood into the aorta and doing no LV unload. Drop to P-2. Advance 1 cm at a time until a ventricular white trace and a pulsatile green current return. Confirm on echo: inlet mid-cavity, ~3.5 cm below the AV.
Hypovolemia / over-diuresis — diastolic suctiondias suction
Impella Suction
160
-40
Ao (mmHg)
98/62(74)
LV
102/-24
835
315
Motor Current (mA)
640/510(575)
10 sec.
  • Position still correct
  • White diastole < 0
  • Green chatters in diastole
Position is still correct: red looks aortic, green still pulses. The tell is white diastole — it crashes below zero in mid-diastole and recovers before the next beat. Flow max/min ratio shrinks (the pump only eats for part of the cycle). Green often chatters in late diastole. The inlet has nothing to eat for part of the beat: bleed, over-diuresis, under-fill, or P that outran venous return. Ease P, give volume (or blood), find the leak. Do not climb P. Repeated diastolic suction is how plasma-free Hb climbs overnight.
Continuous suction — empty or obstructedcont suction
Impella Suction
160
-40
Ao (mmHg)
92/76(81)
LV
92/-43
715
195
Motor Current (mA)
480/430(455)
10 sec.
  • White stays negative
  • Traces overlay
  • Green flattened + chatter
White stays negative the whole cycle. Red and white start to overlay. Green flattens and chatters. This is worse than diastolic suction: completely empty LV, inlet against the wall or the mitral apparatus, thrombus / pannus on the cage, or P-level that has vacuumed the cavity. Drop P now. Echo before the next click — you need to see whether the cavity is small (fill) or the inlet is against tissue (reposition). Volume into a normally sized LV that is obstructed at the cage just raises CVP.
RV failure unmasked — suction with a full right sidedias suction
Impella Suction
160
-40
Ao (mmHg)
78/58(65)
LV
80/-18
790
270
Motor Current (mA)
590/470(530)
10 sec.
  • Position still correct
  • White diastole < 0
  • Green chatters in diastole
The Impella tracing looks like hypovolemia — because the LV is empty. The RV never filled it. You will not see that on the AIC alone. Turn to the bedside: CVP high, PCWP low, PAPi below about 1.0, D-sign on echo, a rising lactate that does not match the “dry” story. Treat the RV (inotrope, pulmonary vasodilator, inhaled agent, RP). More P-level on an empty LV is hemolysis. More fluid into a failing RV is the other way to make it worse. The waveform only told you the inlet is hungry. The right heart told you why.
Tamponade — suction physiology, equalized fillingdias suction
Impella Suction
160
-40
Ao (mmHg)
72/58(63)
LV
76/-14
755
235
Motor Current (mA)
560/430(495)
10 sec.
  • Position still correct
  • White diastole < 0
  • Green chatters in diastole
Again the AIC looks like an empty LV — the cavity cannot fill against pericardial constraint. The giveaway is not on this screen. Equalized RA / PCWP, a blunted y descent, a swinging heart, oliguria after a bleed or post- cardiotomy. Filling the tank does not fix tamponade and will not restore a clean white valley for long. Ease P so the inlet stops chewing wall, then drain. If you only treat the suction alarm you will watch the lactate rise with a “normal” Impella position.
High afterload — motor working, flow notafterload
160
-40
Ao (mmHg)
118/88(98)
LV
122/16
1160
640
Motor Current (mA)
980/820(900)
10 sec.
  • Position correct
  • Green mean high
  • Flow stuck low
Position is fine. White still dives (the inlet has blood). Green is pulsatile but the mean current is high and displayed flow is stuck (P-8, 1.8 L). MAP looks reassuring because MAP is flow × SVR. The leftover norepinephrine from the shock bolus is now a clamp on an axial pump. Wean the pressor or vasodilate. P-9 into a brick wall shears cells and does not unload the LV — white diastolic will start to rise as the ventricle stays loaded. This is the opposite of motor failure: current up, flow down.
No native ejection — LV never crosses Aolow native
160
-20
Ao (mmHg)
67/65(65)
LV
61/-4
1000
0
Motor Current (mA)
590/570(580)
10 sec.
  • White never crosses red
  • AV not opening
  • Native CO ≈ 0
Red is an almost-flat aortic line (here 67/65). White is still ventricular — it peaks and dives — but the LV systolic peak stays below the Ao line (61 versus 67). The left ventricle never generates enough pressure to open the aortic valve, so native cardiac output contributes little to overall cardiac output. Displayed Impella flow (P-9, ~5.5 L on a 5.5) is essentially the cardiac output. Green keeps only a small ripple: there is still a cyclic LV–Ao gradient, but no AV opening to make a full current pulse. Mean current is in the working band — this pump is moving blood. Echo: inlet mid-cavity, outlet above the AV, leaflets closed. Do not pull it. Do not call this malposition (red and white are still different chambers). Do not confuse it with token ECMELLA (white diastolic would be high, not −4).
VT / no fill timevt
Placement Signal Not Reliable
160
-40
Ao (mmHg)
58/46(50)
LV
64/-8
870
350
Motor Current (mA)
700/520(610)
10 sec.
  • Fast irregular spikes
  • Signal not reliable
Fast, irregular spikes. The AIC will often say the placement signal is not reliable — believe it. Every run empties the LV and suction fires, then sinus returns and the numbers recover. Climbing P during a burst suctions a quivering cavity and writes a hemolysis note. Treat the rhythm (amio, shock if unstable, fix K / Mg / ischemia). Come back to P-level after there is a diastole again.
Motor failing, thrombosis, or inlet obstructionmotor fail
160
-40
Ao (mmHg)
70/58(62)
LV
74/18
525
5
Motor Current (mA)
280/250(265)
10 sec.
  • Green mean collapsed
  • Opposite of afterload
P-level unchanged, motor current falls and stays low, displayed flow collapses. That is the opposite of afterload (high current) and the opposite of a healthy seated pump (~500–900 mA depending on P). Think clot in the motor, inlet occluded by tissue or thrombus, a dying purge that let blood into the gap, or a pump that has stopped spinning. White diastolic creeps up — the LV is no longer being unloaded. Echo the inlet, check purge pressure / flow, call for exchange or escalate. Do not wean a pump that has already stopped working, and do not treat this as “needs more P.”
ECMELLA — VA afterload, token Impellaecmella
160
-40
Ao (mmHg)
68/64(65)
LV
70/28
1000
0
Motor Current (mA)
620/580(600)
10 sec.
  • Pulse pressure ~4 mmHg
  • White diastolic high
  • AV not opening
Pulse pressure 4 mmHg. White diastolic is high — the LV is still loaded. Green barely pulses because the AV is not opening. Retrograde VA flow is the clamp; a P-2 Impella next to 4–5 L of VA is a decoration, not an unload. Titrate Impella P until PCWP falls and a pulse returns. If you instead climb into suction, VA is over-draining or the RV is out. Wean leftover vasopressor — more afterload is the opposite of the job.
Impella RP — seated, forwardingrp seated
160
-40
PA (mmHg)
34/18(23)
CVP
14/10
930
410
Motor Current (mA)
710/630(670)
10 sec.
  • PA over CVP
  • Green pulsatile
RP’s placement window is PA (red) over CVP (white), not Ao / LV. Pulsatile motor current still means the inlet and outlet see two chambers (IVC/RA vs PA). A seated RP should drop CVP and raise PA pulse as it forwards the RV. Titrate until CVP falls and the LV Impella (if present) stops sucking. The LV cannot eject what it never receives.
Impella RP — not forwardingrp fail
160
-40
PA (mmHg)
22/16(18)
CVP
22/18
794
274
Motor Current (mA)
540/528(534)
10 sec.
  • PA pulse collapsed
  • Green flat
  • CVP not falling
PA pulse has collapsed onto CVP. Green is flat. The RP is either under-P, malpositioned (outlet not in the PA), or pumping into a pulmonary bed that will not accept flow. CVP stays high; the LV Impella, if there is one, starts to suction. Fix the RP (position, P-level, PE / high PVR) before you blame the left-sided pump. Climbing the LV P-level here is how you hemolyze an empty left ventricle.

Looks like X, is actually Y

Several catastrophes share an Impella picture. The AIC names the inlet. You name the reason.

AIC pictureCould beWhat splits them
Diastolic suctionHypovolemia, bleed, over-diuresisCVP low, small LV, dry exam, falling Hb
Diastolic suctionRV failureCVP high, PCWP low, PAPi < ~1, D-sign
Diastolic suctionTamponadeEqualized filling, blunted y, swinging heart
Diastolic suctionHigh PEEP / abdominal pressureVent / abdomen. Drop PEEP or decompress.
White never crosses redPump-dependent, AV closedHigh pump flow, white diastolic low, echo AV shut
White never crosses redToken ECMELLA, LV still loadedWhite diastolic high, pulse pressure ~4, VA running
Flat green currentIn LV or in AoBoth pressure traces look like the same chamber
Flat green currentNo native ejection, seatedRed and white still different; white stays below red
Flat green currentRP not forwardingYou are on an RP. PA ≈ CVP.
Low flow, high MAPAfterloadGreen mean high, pulse kept, white valley present
Low flow, falling MAPMotor / clot / purge deathGreen mean collapsed, white diastolic rising
Placement not reliableVT, CPR, extreme suctionLook at the patient, not the next P click

What waveforms will not tell you

  • They will not distinguish hypovolemia from RV failure or tamponade. You need CVP, PCWP, PAPi, and echo.
  • They will not give you a true arterial pressure. Use the A-line. AIC “Ao” is an optical sensor at the outlet.
  • They will not diagnose HIT, a cold foot, or a soaking groin. Look at the patient.
  • A flat motor current is not a diagnosis. It is a fork: malposition versus no native beat versus a dead motor.
  • Displayed flow is not cardiac output. Add leftover native stroke volume, and only after position is honest.
  • They will not tell you the inlet is 3.5 cm below the AV. Echo or fluoro does that.

P-levels are not RPM

You do not set revolutions. You set a P-level from P-1 (wean / insertion) to P-9 (max). On a filled CP, P-2 is about 1.6 L/min and P-8 is about 3.3 L/min — afterload and preload then take their cut. 5.5 numbers are higher at the same P; RP numbers live in the CP range but on the right side.

Titrate P only after the placement signal is aortic and echo (or fluoro) has the inlet mid-cavity. Climbing P into suction, malposition, or a brick-wall SVR shears red cells and does not raise effective flow.

P-levelCP (filled)Typical use
P-1 / P-2~1.1–1.7 LInsertion, wean, brief imaging
P-3 – P-5~1.9–2.7 LModest support, recovering LV
P-6 – P-8~2.7–3.3 LSCAI C/D working range on CP
P-9~3.5–4.0 LCeiling. If still shocked, escalate the platform.

Inlet mid-cavity, outlet above the valve

For CP / 5.5 the inlet cage belongs in the mid-LV, about 3.5 cm below the aortic valve. The pigtail sits in the LV. The outlet sits in the aorta, above the AV. Too shallow: outlet in the LV, no forward work, odd signal. Too deep: inlet against the mitral apparatus or the wall, suction and hemolysis.

SmartAssist first, then echo. A ventricular placement signal plus a flat-high motor current is malposition until imaging says otherwise. Reposition under echo or fluoro. Do not climb P on a device sitting in the wrong chamber.

Suction means the inlet is empty

The axial pump can only move what the cavity contains. SUCTION on the AIC is the Impella equivalent of ECMO chatter: the well is dry. Causes, in the order you should think them:

  • Hypovolemia / over-diuresis / bleed
  • RV failure — the LV never fills (watch PAPi and a D-sign)
  • Malposition — inlet against wall or MV
  • Tamponade or high PEEP / abdominal pressure
  • VT / no diastolic fill time

Do not climb P into suction. Ease P-level, restore fill (or treat the RV / rhythm / position), then come back up. Repeated suction is how these devices hemolyze.

High SVR is a clamp

Axial pumps are afterload-sensitive. A P-8 CP into an SVR of 2100 may deliver 1.8 L with a reassuring MAP — because MAP is flow × resistance. The ventricle is not unloaded. Motor current runs high. Wean the leftover norepinephrine or vasodilate. Climbing to P-9 into a brick wall just shears cells.

Purge keeps blood out of the motor

A dextrose purge (D5W ± unfractionated heparin, or a bicarbonate / heparin-free recipe) runs through the motor at a few to a few dozen mL/h. Working purge pressure is typically a few hundred mmHg (about 300–700). The point is to keep blood from entering the motor gap.

A climbing purge pressure with a falling purge flow is a cassette, line, or seal problem. Exchange / prime the purge. It is not a P-level problem. Ignore it and the motor seizes.

Hemolysis is shear

Rising plasma-free Hb, tea-colored urine, falling Hb, drifting motor current. The number is a symptom. The cause is almost always suction, malposition, or high P into a dry or tight ventricle. Fix the inlet. Transfusion alone does not stop the shear.

SCAI shock, CPO, and PAPi

SCAI stages A–E are how adult cardiogenic shock is now spoken. This lab lives in C–E. Two numbers from the cath-lab literature belong on the glass:

CPO (W) = MAP × CO / 451

Fincke et al. (JACC 2004): CPO below about 0.6 W is the shock cutoff associated with mortality. In this lab, CO is approximated as pump flow plus leftover native output (~1.2 L when the LV is wrecked). CPO still under 0.6 W on max CP means escalate the platform.

PAPi = (PASP − PADP) / RA

Korabathina et al. (CCI 2012): PAPi below about 1.0 unmasks an RV that will not feed the Impella. Suction + high CVP + low PCWP + a D-sign is this pattern. Support the RV. More P-level on an empty LV is how you write a hemolysis note.

The groin, the foot, and the cassette heparin

  • Limb ischemia — the 14-Fr CP sheath occupies the femoral artery. Cool foot, weak Doppler: antegrade / distal perfusion sheath now, not after the CK peaks. 5.5 trades groin ischemia for an axillary cutdown.
  • Access bleed — the most common Impella complication. Hold purge heparin, pressure / figure-of-eight, transfuse. Do not explant a working pump because the dressing is red.
  • HIT — cassette heparin is systemic enough to cause and perpetuate HIT. Day-5 thrombocytopenia and mottling: stop every heparin source, switch to bivalirudin (or argatroban), run a heparin-free purge.

Step the P-level down

Wean by stepping P, not by pulling at P-8. Watch MAP, lactate, CPO, and whether the LV dilates on echo. P-2 / P-1 is a wean setting. P-0 is off — do not linger (stasis, thrombus). Have a re-escalate plan. One good gas is not a graduation.

VA-ECMO loads the LV. Impella unloads it.

Femoral VA return is retrograde aortic flow. That afterload can close the aortic valve, smoke the LV, and thrombose a still cavity. An Impella across the AV unloads that LV and lets the valve open — if you actually titrate it. Danial and the ECMELLA series are consistent: token P-2 next to 4–5 L of VA is not an unload.

Titrate Impella P until PCWP falls and the AV opens. The inlet still needs volume. If you climb into suction, VA is over-draining or the RV is out. Wean leftover vasopressor; more afterload is the opposite of the job.

CP, 5.5, and RP are not interchangeable

DeviceAccessJobCeiling
CP14-Fr femoralLV → aorta~3.5–4 L. Escalate if still shocked.
5.5Axillary graftLV → aorta, higher flow / longer support~5.5–6 L. Still needs fill.
RPFemoral veinIVC/RA → PA (RV forward)CVP should fall; LV should fill.
ECMELLAVA + LV ImpellaECMO for flow, Impella for unloadTitrate both. Watch the inlet.

How to work a case in this lab

Same order every time — it is the order that keeps people alive at 03:00.

  1. Look at the patient: MAP, lactate, urine, the foot, the urine color, the groin.
  2. Look at the AIC: P-level vs flow, placement signal, motor current, purge pressure. Compare signal to what an aortic waveform should look like before you touch P.
  3. Name the problem in one sentence (settings too low, suction, malposition, afterload, purge, RV, HIT, token ECMELLA…).
  4. Do the matching move. Harmful clicks treat the wrong organ.
  5. If you neglect shock, suction, or a misplaced inlet, the patient will declare — pre-arrest, then a fail. There is no ROSC mini-game. Restore MAP and flow.

Guided mode forces three decisions. Free play is the same physiology without the multiple choice. The debrief is the attending in the doorway.